Set Descending Direction

5 Item(s)

  1. Light source for Variable Wavelength Detector. Typical replacement interval is every 2,000 working hours Learn More
  2. Replacement rotor seal for collector valve. Replacement interval is every 50,000 switching cycles. Learn More
  3. In-line filter cartridge installed in pump purge valve inlet fitting. Filters eluents between the pump heads and the purge valve (before the mixer). Stainless steel frit, 2 µm pore size, 20 µL volume. Replacement interval is every 5,000 working hours. Learn More
  4. Kit for pump head maintenance: set of gaskets, 2 piston rods, 2 sapphire backing rings, 2 O-rings. Part of annual preventative maintenance. Learn More
  5. Kit for injection valve maintenance; includes normal wear parts, and tools for disassembly/reassembly. Part of annual preventative maintenance. Learn More

Set Descending Direction