Flow Cells
Microwave chemistry has long been known to accelerate chemical transformations and provide a simple means to access more rigorous reaction conditions. Flow chemistry has demonstrated similar benefits, with the added bonus of being able to increase the reaction scale simply by prolonging the reaction. The combination of the two technologies opens the door to a quick optimization and easy scale-up of virtually unlimited types of chemical transformations. The Discover® SP System provides the means to perform both types of reactions simultaneously or sequentially within the same system. Use of microwave heating allowed access to a temperature of 190 °C, well above the boiling point of acetonitrile. This increased temperature resulted in a pressure of nearly 200 psi during the batch reaction. Despite these forcing conditions, the reaction proved suitable for synthesis in flow, providing a modest increase in yield with a 20-fold increase in scale.
10-mL and 80-mL Flow Cells
Explore continuous flow chemistry in the microwave using two standard flow cell sizes.MAX. TEMPERATURE: 300 °C
MAX. PRESSURE: 18 bar (250 psi)
View Flow TechNote
Table 1: Heck Reagents Required
| Reagent | Batch | Flow |
| Iodobenzene | 0.101 g | 2.02 g |
| Butyl acrylate | 0.127 g | 2.54 g |
| Pd(OAc)2 | 0.06 mg | 1.2 mg |
| DIPEA | 0.096 g | 1.92 mg |
| MeCN | 1 mL | 20 mL |
Table 2: Heck Coupling Conditions
| Mode | Temperature | Time | Scale | Isolated Yield |
| Batch | 190 °C | 5 minutes | 0.5 mmol | 89% |
| Flow | 190 °C | 30 minutes | 10 mmol | 91% |

Pressure Tube Fitting with Ferrule, 0.25" Extension
Part # 165990Pressure tube fitting with ferrule and 0.25" extended tube for the 10 mL Discover flow cell.
Saber más
Pressure Tube Fitting with Ferrule, 3.65" Extension
Part # 165995Pressure tube fitting with ferrule and 3.65" extended tube for the 10 mL Discover flow cell.
Saber más